TECHNICAL DATA YASHAM 635
Patient input circuit: Fully floating and isolated, defibrillation protected
Monitor display
12-channel display or Rhythm
Heart Rate
2.5,5,10,20 mm/mV or Automatic in Auto Mode
Filter status (Notch , Low-pass , Drift)
Insufficient electrode contact(Lead-off)
Calibration voltage : 1mV±3%
Leads: 12 simultaneous Standard leads
Chart Printout speed: 6.25/12.5/25/50 mm/s
Sensitivities 2.5/510/20 mm/mV, either automatically adjusted or manually selected
Automatic lead programs
1/3/6-channel presentations of 12 simultaneously recorded leads
1+long lead/3+long Lead/Rhythm
Data record
Patient data (name, age, height, weight,(User ID
Listing of all ECG recording conditions (data, time, filter)
Measurement program :ECG measurement results(intervals, amplitudes, electrical axes)
Average complexes with measurement markings
Recording track: 1/2/3/4/6-channel presentations optimal positioning on a width of 112 mm Automatic baseline adjustment
Filter: Myogram filter: 25 or 35 or 75 Hz or off, Drift: baseline stabilizer
ECG amplifier
Simultaneous recording of all 9 active electrode signals (=12 leads)
Sampling frequency: 1000 Hz
Pacemaker Detection/Rejection:±2~±700mV/0.1ms~2ms
System Dimensions: 350*284*65mm,approx.2.0 kg
Built-in monitor: 95*53 mm effective display area ,480*272 dots resolution
On-screen status indicators: power source ,Battery status, date, time
Control panel and key board: User-friendly keys ,LED indicator, Touchscreen
Power supply requirements: 90-240V,50/60Hz,stand-alone operation with Built-in rechargeable Li-ion battery, integrated power supply unit
Battery capacity: 5 hours of normal use
Power consumption: Max 30 VA
Frequency range of digital recording system: 0.05-150Hz(IEC/AHA)
Chart paper: Thermo-sensitive, 112 mm wide
Printing process: High-resolution thermal head printer ,8 dots/mm
Interfaces: USB Interfaces ,data transmission to PC, optional Ethernet interface
Memory: Possibility to save up to 100 Resting ECG measurements internally or SD card (optional)
Environmental conditions:
Temperature, operating: 10°C to 40°C
Temperature, storage: -10°C to 50°C
Relative humidity: 25% to 95% (no condensation)
Pressure during operation: 700 to 1060 hpa
Standards:
Safety: IEC(EN)60601-1,IEC(EN)60601-1-2(EMC)
Performance: IEC(EN)60601-2-25,IEC(EN)60601-2-51,ANSI/AAMI EC11
Protection class: I according to IEC 60601-1(with internal power supply)
Applied part: CF according to IEC 60601-1