TECHNICAL DATA YASHAM 310

logo
TECHNICAL DATA YASHAM 310

 

TECHNICAL DATA YASHAM310 -2

Technical Data Yasham310 

Technical Data for ECG 

Patient input circuit: fully floating and isolated, defibrillation protected 

Monitor display
   12-channel display or Rhythm 
   Heart Rate 
   2.5,5,10,20 mm/mV or Automatic in Auto/Manual Mode
   Filter status (Notch, Low-pass,Drift,EMG) 
   Insufficient electrode contact (Lead- Off)

 Leads:12 simultaneous Standard leads 

Chart Printout speed: 6.25/12.5/25/50 mm/s 

Sensitivities: 2.5/5/10/20 mm/mV, either automatically adjusted or manually selected 

Automatic or Manual lead programs: 
   1/2/3- channel presentations of 12 Simultaneously or Real time recorded leads 
   1+long Lead/2+long Lead/Rhythm 

Data record 
   Patient data (name, age, height, weight, stature, weight, sex, blood group & heart rate), User ID 
   Listing of all ECG recording conditions (data, time, filter, speed) 

Measurement program: ECG measurement results (intervals, amplitudes, electrical axes) average complexes with measurement markings 

Calibration voltage: 1mV3% 

Recording track: 1 / 1+1 /2 /1+2/ 3-channel + Rhythm Mod 

Filter: Myogram filter:25 or 35 or 75 or150 Hz or Off, Drift: baseline stabilizer, Notch:50/60 Hz 

ECG amplifier: Simultaneous recording of all 9 active electrode signals (=12 leads) 

Sampling frequency: 1000 Hz

 Pacemaker Detection/Rejection:.1ms 

System Dimensions: 210*260*65 mm, approx .1.8 kg 

Built-in monitor: 65*110 mm effective display aree; 480*800 dots resolution

 On-screen status indicators: power source, Battery status, time 

Control panel and key board: User-friendly keys, LED indicator, Touchscreen

 Power supply requirements: 90- 264 V , 50/60Hz ; stand- alone operation with Built-in rechargeable Li-ion battery, integrated power supply unit 

Battery capacity: 8 hours of normal use

 Power consumption: 90 

Frequency range of digital recording system: 0.05 – 150 Hz (IEC/AHA) 

Chart paper: Thermo-sensitive, 63 mm wide

 Printing process: High-resolution thermal head printer, 8 dots /mm 

Interfaces: USB Interfaces; data transmission to PC; optional Internet interface

 Memory: Possibility to save up to 100 Resting ECG measurements internally or SD card (optional) 

Environmental conditions 
   Temperature, operating: 10 C to 40C 
   Temperature, storage: -10 C to 50C 
   Relative humidity: 25% to 95% ( no condensaton) 
   Pressure during operation: 700 to 1060 hpa

 Standards 
   Safety: IEC(EN)60601-1; IEC(EN)60601-1-2(EMC) 
   Performance: ANSI/AAMI EC11 
   Perotection class: I according to IEC 60601-1 (with internal power supply) 
   Applied Part: CF according to IEC 60601-1 
   Hardware options: Equipment trolley 

Optional 
   USB HOST 
   USB Device 
   LAN(RJ-45)
   Wifi 
   SD Card

 